CyberRota Analysis
This is a medium severity vulnerability with a CVSS score of 5.9. See the original NVD description below for full technical details.
CVE
CVE-2017-12721
Severity
MEDIUM
CVSS
5.9
EPSS
0.67%
Original NVD Description
An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.